Oferta!

Bomba de infusão de seringa médica THR-IP100

Descrição

Bomba de infusão de seringa médica THR-IP100

THR-IP100 Auto Caliration Infusion Pump

Material:Plástico ABS,Fire-proof
Dimensão135L*120W*245H mm
Peso:2kg
Special Features
Precise infusion rate control ml/h and d/min Switch Display Interface
Self-diagnosis, infusion pipe installation ,pressure setting,rate setting.English LCD interface
KVO function starts once infusion completed. Night Mode
Intellectualized drop detector Hard based Free-flow protection
Alarm Function Conditions:
Infusion completeOcclusion
Low batteryInstallation error
Air-on-lineAbnormal rate
Door openOperational error
Screen Display Context
Battery charge indicatorOcclusion value: High,Middle and low
Accumulated ValueInfusion complete
AC power indicatorPower capacity
Preset Infusion ValueInfusion speed
Fast infusion rateInfusion status
Alarm type
Especificação
Pump Mechanism: Peristalsis Mechanism
Work Mode: -Continuous working mode
-Control ModeDrop
-Detector Mode
Rate Range:Preset flow rate range: 1-250d/min or 1-1000ml/h.
( 1d/min or 1ml/h increment)
Preset Volume Limit: 5-9999 ml
Accumulated Volume: 0-9999 ml
Precisão: ±5%
KVO Rate(Keep vein open): 1ml/h
Occlusion Value Levels: Adjustable from High, Middle and low level
AC Power Supply:AC 85–264V, 50/60 Hz
Tipo de Bateria:DC12V,1800mA/h NI-MH rechargeable battery
Bateria : Can use for more than 3 hours after fully recharge(run at middle rate)
Battery Recharge: When connect to AC power, it automatically Recharge and use about 15 hours to fully recharge
Consumption:<28VA
Fuse:F 0.8A H/250V
Security Level:Class 2,class BF, internal battery, water-proof
Operation environment :Environment temperature +5°C+40°
Pressão atmosférica:86-106kPa
Relative humidity:20%~90%
OR
Environment :Environment temperature -22°C-55°C
Pressão atmosférica:50~106kPa
Relative humidity:90%
Electrical Safety:In accord with the stipulation of EC 60601-1:1998And EN 60601-1
Características:
Hard based free-flow protection:When the door open ,it effectively prevents any free-flow .Prevent any danger to the patients.
Drop detector:
It’s used to calculate the drop infused for the patients, Meanwhile it automatically adjust the accuracy by itself. According to the current infusion set. One of another, the detector can also detect the abnormal flow rate and sends this information to the CPU, The CPU sends instruction to control the peristalsis pieces, accordingly make the accurate infusion come true.
Night mode:

At night, in order to ensure the patients have a good rest, this pump has the silence function and screen-protection mode.

Embalagem & Envio

Informações da Empresa

Nossos serviços

Informação adicional

Tipo:

Infusion Pump with Auto Caliration

Marca:

THR

Número do modelo:

THR-IP100

Lugar de origem:

Jiangsu, China

Classificação do instrumento:

Classe I

garantia:

1 Ano

Serviço pós-venda:

Suporte técnico on-line

Nome do Produto:

Caliration Infusion Pump

AC power supply:

CA 100 to 240V±10%,50/60Hz;

MOQ:

1 unit

Certificado:

CE ISO

OEM:

Disponível